Experimental tests



A number of experiments have been carried out to test the accuracy and resolution of the instrument applied to:

Model systems

The sizes of calibration latexes with mean radii 0.05, 0.1, 0.135, and 0.25 microns dissolved in water and with the width of the polydispersity distribution less than 6% have been measured. The latexes were produced and calibrated by the Lebedev Rubber Research Institute, St. Petersburg, Russia. With the assumption that the polydispersity distribution is symmetrical, the inaccuracy of the mean radius determination does not exceed 1.5 % in each case.



Opaque systems

The particle-size measurements in an opaque system (black aniline dye dissolved in water) have also been performed. The dye concentration was changed from that corresponding to absolutely transparent to opaque solutions. In the latter case the light penetration into the sample did not exceed 0.5 mm. As is seen from the figure below, for all dye concentrations, the measured radius is stable within 5%.
[Picture]

The measurements were carried out with ”backscattering” 1350 geometry (solid squares in the above figure). The conventional 900 geometry in a cylindrical cell has also been used for the smallest concentration (0.01%) at which the solution is transparent(a circle in the figure).



Dilute polymer solutions

The diffusion coefficient and particle size of polystyrene of molecular weight Mw = 96,400 dissolved in toluene, at 22 0C has been measured. The mass concentration of polystyrene was 0.0217 g cm-3. Two optical geometries and sample cells were used: with scattering angles of 450 (a round glass cell of the diameter 27.25 mm) and 900 (a square cell of 12.5 x 12.5 mm). The scattering in this system is very weak. Such conditions are not favorable for accurate measurements. Nevertheless, excellent agreement with precision other-technique measurements of the diffusion coefficient in this system has been obtained. See Refs. [19,20] in the Related literature section.



Back to the Photon-Correlation Spectrometer page